_edited.png)
GTI 3991
Aerosol Photometer
Check out our Catalog Pdf.
Applications:
Independent filter certifiers
Nuclear facilities
Pharmaceutical manufacturers
Medical facilities and cleanrooms
Biosafety cabinets and fume hoods
Key Features & Benefits:
8.4inch color touch screen (English & Chinese)
Built in thermal printer and equipped with
thermal printing paper for real-time printing
The scanning probe has one-dimensional code scanning function,
automatic data classification, and traceability
Large storage capacity, capable of storing 100000 sets of data
Built in lithium battery, which lasts for 4.5 hours
Built in GMP Three-Level Privileged Management which
is able to generate operation logs
Built in low-power mode to reduce instrument power consumption without testing
Performance & Product Specifications:
Data Accuracy
| 0.0001~100.0% 8.4inch Touch LCD Screen |
Dynamic Range
| 0.0001µg/L~600µg/L |
Accuracy
| 1% of Readings(0.01~100%)
|
Repeatability
| 0.5% of Readings (0.01~100%)
|
Flow Control
| 1cfm (28.3L/min)±5% |
Zero Building
| Automatic or manually |
Alarm Modes
| Audible, visual and vibratory
|
Power-Saving Mode | Lower energy consumption |
Reporting Modes
| Continuous, Summary, Monitoring mode
|
| Built-in thermal printer
|
ID Encode
| ID scanning, data classify and traceable.
|
Data Output
| Storage, USB, Print; Logs can be traced and stored |
Data Storage Capacity | 100,000 Groups |
Communication | USB is connected with PC continuous scanning testing is available. |
GMP Three-Level Privileged Management | Contains three-level permission management, which can set identities and assign usage permissions; it can be done with Log Report |
Language | English |
Reagent Settings
| PAO、DOP、Ondina、PolvethleneGlvcol.Corn Oil、MineralOil、Paraffin Oi, etc. |
Sampling Port | Standard internal rectangular sampling port without unnecessary corners |
Shell Material | SUS304 stainless steel |
Temp. & Humidity | 5~50°C (No condensation, no freezing); 5~95%RH |
Storage Temp. | -20C~60°C (Relative humidity below 95%, no condensation) |
Power
| AC100~240V 50/60Hz or lithium battery (optional with built-in lithium battery pack) |
Dimension/Weight
| Approx.220x260 (handle 300) x 390mm (W x H x L) 11.1kg (photometer) probe and accessories 1kg |
Reference complied
| NSF49、IEST、ISO14644、EN61010-1:2010、EN61326:2006 |
What’s included | Photometer, scanning probe, sampling air pipe, manual, certification, power cord, carrying case |
Accessories | Lithium battery (built-in), AC/DC, circular sampling port, square sampling port |